Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25459 |
| Brand Name: | Cios Alpha / Fusion |
| Model: | C-arm holder for all Cios Alpha systems |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare GmbH |
| Issue: | During a routine inspection of a Cios Alpha system, a crack on the C-arm holder was discovered. In the event the C-arm holder has a crack, the stability of the C-arm may be compromised and could lead to deformation of the C-arm. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 10/12/2019 |