Detail of Recall

Type of Product:Device
Medsafe Reference:25454
Brand Name:MRI Systems (ONI)
Model:Discovery MR750w 3.0T, Discovery MR750 3.0T, Optima MR450w 1.5T, Discovery MR450 1.5T, SIGNA Architect, SIGNA Artist and SIGNA PET/MR
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:ONI Medical Systems Inc
Issue:Certain Main Disconnect Panels (MDPs) from various third party suppliers may not meet the manufacturer preinstallation specifications. As a result, when the Emergency Off (E-Off) circuit is activated, the MDP will not remove power to the Heat Exchanger Cabinet (HEC). Although this condition by itself does not pose a risk, the forced air from the cooling fan in the HEC may accelerate an unrelated thermal event (e.g., fire or smoke).
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/12/2019