Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25428 |
| Brand Name: | ARCHITECT Reaction Vessel (RV) |
| Model: | |
| Affected: | ARCHITECT Reaction Vessels - List number 7C15-03 |
| Software version: | |
| Recalling Organisation: | Abbott Diagnostics Division Ground Floor, Building D 4 Pacific Rise Mt Wellington AUCKLAND 1640 |
| Contact Information: | 0800 656 233 |
| Manufacturer: | Abbott Laboratories (NZ) Ltd |
| Issue: | The manufacturer has observed an increase in complaints due to static discharge in the ARCHITECT Reaction Vessel (RV) on the ARCHITECT i2000, i2000SR, and i1000SR at the optics reader. The static discharge can result in Error Code 1007 (Unable to process test, activated read failure) or potentially falsely elevated or false positive patient results when using the ARCHITECT RV lots |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 3/12/2019 |