Detail of Recall

Type of Product:Device
Medsafe Reference:25428
Brand Name: ARCHITECT Reaction Vessel (RV)
Model:
Affected:ARCHITECT Reaction Vessels - List number 7C15-03
Software version:
Recalling Organisation:Abbott Diagnostics Division
Ground Floor, Building D
4 Pacific Rise
Mt Wellington
AUCKLAND 1640
Contact Information:0800 656 233
Manufacturer:Abbott Laboratories (NZ) Ltd
Issue:The manufacturer has observed an increase in complaints due to static discharge in the ARCHITECT Reaction Vessel (RV) on the ARCHITECT i2000, i2000SR, and i1000SR at the optics reader. The static discharge can result in Error Code 1007 (Unable to process test, activated read failure) or potentially falsely elevated or false positive patient results when using the ARCHITECT RV lots
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/12/2019