Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25406
Brand Name:Elecsys Troponin
Model:
Affected:Elecsys® Troponin T hs V2 (300 tests) and Elecsys® Troponin T hs (300 tests), Lots - 419260, 436777, 474941, Elecsys® Troponin T hs (300 tests), Lot - 429178
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:The manufacturer has received complaints regarding non-reproducible elevated results (highflyers) for specific reagent lots of Elecsys® Troponin T hs on cobas e 801. This updated Recall for Product Correction is to inform laboratories that Elecsys® Troponin T hs product code 08469873190 lot 474941can be used without restrictions and as such double determinations are no longer required for this reagent lot. Reagent lot 474941 is the current lot number in New Zealand stock.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/6/2020
Hide menus