Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25383 |
| Brand Name: | Xper IMS |
| Model: | |
| Affected: | Xper Information Management System version 2.4.2 and 2.4.3 |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems Nederland BV |
| Issue: | The manufacturer has become aware of a potential problem with our Xper Information Management System version 2.4.2 and 2.4.3. The problem was detected regarding the calculations of PVR, TPVR, SVR and TSVR. Due to an error in the software code CO is used for all PVR and SVR calculations when using the Fick principle. In a heart with no shunt the calculation result will be correct as QS,Qp and CO are normally equal. In the presence of a shunt the calculated resistance value is incorrect. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 19/11/2019 |