Detail of Recall

Type of Product:Device
Medsafe Reference:25383
Brand Name:Xper IMS
Model:
Affected: Xper Information Management System version 2.4.2 and 2.4.3
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:The manufacturer has become aware of a potential problem with our Xper Information Management System version 2.4.2 and 2.4.3. The problem was detected regarding the calculations of PVR, TPVR, SVR and TSVR. Due to an error in the software code CO is used for all PVR and SVR calculations when using the Fick principle. In a heart with no shunt the calculation result will be correct as QS,Qp and CO are normally equal. In the presence of a shunt the calculated resistance value is incorrect.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/11/2019