Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25368 |
| Brand Name: | Tornier Aequalis Ascend Flex |
| Model: | Wright Medical Aequalis Ascend Flex Humeral Stems |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Paragon Care Group New Zealand Limited 69 Elizabeth Knox Place Glen Innes AUCKLAND |
| Contact Information: | 0508 654 258 |
| Manufacturer: | Tornier SAS |
| Issue: | The manufacturer, has advised that during shipping, it is possible for the Aequalis Ascend Flex humeral stem to become unseated or otherwise move within its internal packaging restraints. Movement within the packaging may cause the product to rub against the PETG blister shipping container causing the transfer of PETG material from the blister container to the stem. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/11/2019 |