Detail of Recall

Type of Product:Device
Medsafe Reference:25363
Brand Name: ARCHITECT CRP Vario and Alinity c CRP Vario Reagent Kits
Model:
Affected:Multiple Lots
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:Abbott Australasia Pty Ltd (NZ)
Issue:Internal studies were carried out by the manufacturer of the ARCHITECT CRP Vario and Alinity c CRP Vario reagents, to verify the sample stability claims in the current Instructions for Use. The manufacturer has confirmed that the EDTA specimen collection tube type (Dipotassium and Tripotassium) did not meet the sample storage stability claim listed in the package insert and is no longer appropriate for use.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/11/2019