| Type of Product: | Device |
| Medsafe Reference: | 25346 |
| Brand Name: | Muzuho OSI Levo System |
| Model: | |
| Affected: | Levo Arm - 7887-050 and 7887-050R |
| Software version: | |
| Recalling Organisation: | Lifehealthcare Ltd New Zealand Unit E 61 Hugo Johnston Drive Penrose Auckland 1642 |
| Contact Information: | |
| Manufacturer: | Mizuho OSI |
| Issue: | The manufacturer has became aware of the potential for fluid ingress resulting from unapproved cleaning techniques. Examples of incorrect cleaning could include submersion or exposure of the unit to direct fluid flow. This may result in fluid entering the unit, rendering the internal electronics inoperable and leading to a failure of the Arm's locking mechanism in isolated situations |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/11/2019 |