Detail of Recall

Type of Product:Device
Medsafe Reference:25346
Brand Name:Muzuho OSI Levo System
Model:
Affected:Levo Arm - 7887-050 and 7887-050R
Software version:
Recalling Organisation:Lifehealthcare Ltd New Zealand
Unit E 61 Hugo Johnston Drive
Penrose
Auckland 1642
Contact Information:
Manufacturer:Mizuho OSI
Issue:The manufacturer has became aware of the potential for fluid ingress resulting from unapproved cleaning techniques. Examples of incorrect cleaning could include submersion or exposure of the unit to direct fluid flow. This may result in fluid entering the unit, rendering the internal electronics inoperable and leading to a failure of the Arm's locking mechanism in isolated situations
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/11/2019