| Type of Product: | Device |
| Medsafe Reference: | 25324 |
| Brand Name: | Infusion pump, syringe |
| Model: | Syramed µSP6000 Syringe Drivers |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Admedus (NZ) Limited c/- BDO Wellington Limited (Accountants) Level 1, 50 Customhouse Quay Wellington 6011 |
| Contact Information: | |
| Manufacturer: | Arcomed AG |
| Brand Name: | Infusion pump, general-purpose |
| Model: | Volumed µVP7000 Large Volume Infusion pumps |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Admedus (NZ) Limited c/- BDO Wellington Limited (Accountants) Level 1, 50 Customhouse Quay Wellington 6011 |
| Contact Information: | |
| Manufacturer: | Arcomed AG |
| Issue: | Investigation revealed that one capacitor on the main PCB was assembled with reverse polarity. Since the capacitor is used more than ten times below its rated voltage, the capacitor operates as desired, but may age faster than expected. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 1/11/2019 |