Detail of Recall

Type of Product:Device
Medsafe Reference:25324
Brand Name:Infusion pump, syringe
Model:Syramed µSP6000 Syringe Drivers
Affected:
Software version:
Recalling Organisation:Admedus (NZ) Limited
c/- BDO Wellington Limited (Accountants)
Level 1, 50 Customhouse Quay
Wellington 6011
Contact Information:
Manufacturer:Arcomed AG
Brand Name:Infusion pump, general-purpose
Model:Volumed µVP7000 Large Volume Infusion pumps
Affected:
Software version:
Recalling Organisation:Admedus (NZ) Limited
c/- BDO Wellington Limited (Accountants)
Level 1, 50 Customhouse Quay
Wellington 6011
Contact Information:
Manufacturer:Arcomed AG
Issue:Investigation revealed that one capacitor on the main PCB was assembled with reverse polarity. Since the capacitor is used more than ten times below its rated voltage, the capacitor operates as desired, but may age faster than expected.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/11/2019