Detail of Recall

Type of Product:Device
Medsafe Reference:25322
Brand Name:External Defibrillator (AED)
Model:HeartStart XL+ Defibrillator/Monitor (Model #:861290),
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The manufacturer has identified that the HeartStart XL+ Defibrillator/Monitor (Model number 861290) may fail to turn on or unexpectedly attempt to restart, rendering it unable to return to a ready for use state. Although the device indicates to the user that it is not ready for use, a failed restart may result in a delay in therapy if the defibrillator/monitor is needed for immediate use.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/11/2019