Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25320
Brand Name:Bard Australia Pty Limited - Biopsy gun, side-notch
Model:New affected lot - REDS1470
Affected:Lot numbers - RECN0753, RECP1763, RECN0756, REDN2448, REDN3426, REDP3393, RECZ2179, RECZ3178, REDR0747, REDN2401, RECN0654,RECN1633, RECQ1208, RECZ3167, REDN0295, REDN2445, REDP5040, REDP2160, REDR0677, REDP3392,REDR2041,REDP4638, REDR3628, REDQ2498
Software version:
Recalling Organisation:Pharmacy Retailing (NZ) Ltd t/a Healthcare Logistics
54 Carbine Road
Mt Wellington
AUCKLAND
Contact Information:09 969 0705
Manufacturer:Bard Peripheral Vascular Inc
Issue:The manufacturer has identified 3 failure modes for the Max-Core® Disposable Core Biopsy Instrument - Failure to fire, firing problem and failure to obtain sample. These issues are associated with one of the molds.
Latest update:
The manufacturer has supplied a recall affected lot (REDS1470) to NZ in error, which they since sold to the market, therefore they need to extend the recall to this new lot.



Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/10/2019