Detail of Recall

Type of Product:Device
Medsafe Reference:25317
Brand Name:Bio-Rad IH-1000
Model:
Affected:Software version: V04.07.03
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Issue:An anomaly of the instrument has been observed by the manufacturer. The IH-1000 shows an incorrect behaviour regarding the transfer of ID-Cards between the pipetting area, the incubators and the centrifuges of the instrument and may perform the assay with a significantly reduced time of incubation at 37°C. The results are returned without flag.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/10/2019