Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25317 |
| Brand Name: | Bio-Rad IH-1000 |
| Model: | |
| Affected: | Software version: V04.07.03 |
| Software version: | |
| Recalling Organisation: | Bio-Rad New Zealand Level 30 48 Shortland Street AUCKLAND 1010 |
| Contact Information: | 0508 805 500 |
| Manufacturer: | Bio-Rad Laboratories Inc |
| Issue: | An anomaly of the instrument has been observed by the manufacturer. The IH-1000 shows an incorrect behaviour regarding the transfer of ID-Cards between the pipetting area, the incubators and the centrifuges of the instrument and may perform the assay with a significantly reduced time of incubation at 37°C. The results are returned without flag. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 30/10/2019 |