Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25300 |
| Brand Name: | Alinity c Carbon Dioxide Reagent Kit |
| Model: | |
| Affected: | lot number 54449UQ03. lot number 54454UQ03. lot number 55731UQ09 |
| Software version: | |
| Recalling Organisation: | Abbott Australasia Pty Ltd (NZ) Abbott Laboratories NZ 4 Pacific Rise, Mt Wellington Auckland 1060 |
| Contact Information: | 61 2 93849717 |
| Manufacturer: | Abbott Australasia Pty Ltd (NZ) |
| Issue: | Internal studies have identified that atmospheric carbon dioxide can be absorbed into the reagent resulting in the potential for incorrect results. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/10/2019 |