Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25300
Brand Name:Alinity c Carbon Dioxide Reagent Kit
Model:
Affected:lot number 54449UQ03. lot number 54454UQ03. lot number 55731UQ09
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:Abbott Australasia Pty Ltd (NZ)
Issue:Internal studies have identified that atmospheric carbon dioxide can be absorbed into the reagent resulting in the potential for incorrect results.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/10/2019
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /