Detail of Recall

Type of Product:Device
Medsafe Reference:25287
Brand Name:Haemonetics Autotransfusion Blood Cell Wash Sets (Class IIb)
Model:Model: CSE-P-125, CSE-P-225, CSE-FP-125V, CSE-FP-225V, 00260-00, 00261-00, 00263-00, and 00265-00 - All Lot numbers
Affected:
Software version:
Recalling Organisation:Haemonetics New Zealand Limited
No 120, 300 Richmond Road
Grey Lynn
Auckland 1021
Contact Information:
Manufacturer:Haemonetics Corporation
Issue:The manufacturer has recently identified a potential issue with Cell Saver® 5/5+ and Cell Saver® Elite®/Elite® + 125ml and 225ml bowl sets. The investigation has determined that a small number of bowl sets may have the potential to develop leaks at the inner core, which could result in fluid becoming trapped inside the bowl core, causing a "long empty" device error message with the possibility of incomplete washing.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/10/2019