Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25287 |
| Brand Name: | Haemonetics Autotransfusion Blood Cell Wash Sets (Class IIb) |
| Model: | Model: CSE-P-125, CSE-P-225, CSE-FP-125V, CSE-FP-225V, 00260-00, 00261-00, 00263-00, and 00265-00 - All Lot numbers |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Haemonetics New Zealand Limited No 120, 300 Richmond Road Grey Lynn Auckland 1021 |
| Contact Information: | |
| Manufacturer: | Haemonetics Corporation |
| Issue: | The manufacturer has recently identified a potential issue with Cell Saver® 5/5+ and Cell Saver® Elite®/Elite® + 125ml and 225ml bowl sets. The investigation has determined that a small number of bowl sets may have the potential to develop leaks at the inner core, which could result in fluid becoming trapped inside the bowl core, causing a "long empty" device error message with the possibility of incomplete washing. |
| Recall action type: | Alert |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 24/10/2019 |