Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25278 |
| Brand Name: | CUSA Clarity : console & footswitch |
| Model: | CUSA Clarity Console (C7000) - all serial numbers |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Integra Neurosciences Pty Ltd (NZ) 12 Florence Avenue Orewa Auckland 0931 |
| Contact Information: | 0800 400 660 |
| Manufacturer: | Integra Lifesciences (Ireland) Ltd |
| Issue: | The manufacturer has discovered that the Operator's Manual needed revision to reflect the most accurate sterilization parameters. The dry time parameters have been changed from 30 minutes to 40 minutes for the CUSA Clarity 36kHz handpiece and components, consistent with the CUSA Clarity 23kHz handpiece and components. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 23/10/2019 |