Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25260
Brand Name:Xen Glaucoma Treatment System(3 models)
Model:XEN 45 Glaucoma Treatment System - Glaucoma shunt
Affected:
Software version:
Recalling Organisation:Allergan New Zealand Ltd
Cnr Manu Tapu Drive & Joseph Hammond Place
Auckland 2022
Contact Information:
Manufacturer:Allergan Inc
Issue:During routine in-process manufacturing inspection of the needle, trace amounts of polishing compounds were observed on the needle of an unreleased XEN® unit. Trace amounts of polishing compounds on the XEN® injector needle could transfer to the patient's eye during procedure possibly resulting in irritation, inflammation, local allergic reaction/ hypersensitivity, iritis, uveitis/sterile endophthalmitis or an intraocular foreign body.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/10/2019