Detail of Recall

Type of Product:Device
Medsafe Reference:25248
Brand Name:GE Incubator, infant, stationary
Model:Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:The manufacturer has identied that if a bedside panel or porthole is not latched will fall open, and it will no longer protect the patient from falling.
Update:
Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation and Giraffe OmniBed Carestation - Bedside panels and portholes can appear closed without being latched - Potential patient fall.
This is a supplement to a previous notification you received and provides additional Safety instructions and materials.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/11/2019