Detail of Recall

Type of Product:Device
Medsafe Reference:25229
Brand Name:Sterile Laryngyscopes Blade
Model:
Affected:BriteBlade Pro Single Use Fibre Optic Macintosh Blade Size 3 - Lot number: 181000637
Software version:
Recalling Organisation:Flexicare Australia Pty Ltd
61 Reid Road
Glenbrook
Waiuku 2681
Contact Information:
Manufacturer:Flexicare Medical Ltd
Issue:The potential for the spring-loaded bearings that secure the blade in position when engaged on the laryngoscope handle to become separated from the blade block and possibly fall into the oral cavity if the blade is disengaged over the patient's open mouth.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/10/2019