Detail of Recall

Type of Product:Device
Medsafe Reference:25210
Brand Name: Emotion 16
Model:SOMATOM go.Top with software syngo CT VA20A_SP2 and with active "Guide&GO" license
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Shanghai Medical Equipment Ltd
Issue:The manufacturer has identified that when adjusting the slice thickness for the i-Sequence to a 3x3 mm colimation in any Guide&GO intervention protocol, the dose (CTDIvoland DLP) is incorrectly calculated as 0.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/10/2019