Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25194 |
| Brand Name: | V60 Ventilator |
| Model: | |
| Affected: | Serial Number - 100256990, 100257691 |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Respironics California LLC |
| Issue: | The manufacturer has become aware of a potential problem with the V60 ventilator where the touchscreen may become frozen and fail to respond to touch commands. Consequently, if a patient requires a change in a ventilator therapy setting, e.g., FiO2, the change could not be made. Although it would be immediately apparent to the user that the changes cannot be made, there is no advance warning of touchscreen failure. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 4/10/2019 |