Detail of Recall

Type of Product:Device
Medsafe Reference:25173
Brand Name:Cardiac pipettes 150 µl 20pc - cobas h232
Model:
Affected:Product code - 11622889190
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:The manufacturer has identified that if the needle was not completely fixed to the body of the pipette and is used for samples transfer from the tube, the needle may become stuck in the cap of the primary sample tube when being pulled out of the tubes. The cap is made from more resistant and elastic material than human tissue, so it depends also on the handling, how the needle is retracted.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/10/2019