Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25173 |
| Brand Name: | Cardiac pipettes 150 µl 20pc - cobas h232 |
| Model: | |
| Affected: | Product code - 11622889190 |
| Software version: | |
| Recalling Organisation: | Roche Diagnostics (NZ) Ltd 15 Rakino Way Mt Wellington AUCKLAND 1644 |
| Contact Information: | 0800 363 622 |
| Manufacturer: | Roche Diagnostics GmbH |
| Issue: | The manufacturer has identified that if the needle was not completely fixed to the body of the pipette and is used for samples transfer from the tube, the needle may become stuck in the cap of the primary sample tube when being pulled out of the tubes. The cap is made from more resistant and elastic material than human tissue, so it depends also on the handling, how the needle is retracted. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/10/2019 |