Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 25157 |
| Brand Name: | Infinity Acute Care System Monitoring - M540 |
| Model: | |
| Affected: | software version VG4.1.1/VG4.0.3 and lower |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draegerwerk AG & Co KGaA |
| Brand Name: | INFINITY Patient Monitors & Accessories |
| Model: | |
| Affected: | software version VG4.1.1/VG4.0.3 and lower |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draeger Medical Systems Inc |
| Issue: | The manufacturer has identified a potential cybersecurity issue which may affect our Infinity Acute Care Monitoring System (IACS) and the Infinity M540 Monitor Stand Alone Monitor; software versions VG4.1.1 & VG4.0.3 and Lower. These vulnerabilities may cause the device to reboot, lose alarm functionality, and/or lose communication with the cockpit and/or the Infinity Network. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 30/9/2019 |