Type of Product:Device
Medsafe Reference:25153
Brand Name:MOSAIQ
Model:
Affected:MOSAIQ Version 2.0 and higher
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Impac Medical Systems Inc
Issue:The user may inadvertently enter Metric values into Height and Weight fields labeled with US Standard units. If the units are left defaulted to US Standard, but Metric measurements are entered, the resulting BSA will be much larger than expected.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/9/2019