Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25149
Brand Name:COBAS® AmpliPrep/COBAS® TaqMan® HCV Test, v2.0
Model:
Affected:COBAS® AmpliPrep/COBAS® TaqMan®
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:The manufacturer has observed abnormal curve shapes that could lead to high titre false positive results when using either the COBAS® AmpliPrep/COBAS® TaqMan® (CAP/CTM) HC Quantitative or Qualitative, v2.0 tests. Internal investigations have found that this issue was caused by early dips” in the PCR growth curves.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/9/2019