Detail of Recall

Type of Product:Device
Medsafe Reference:25075
Brand Name:Oesophageal Transducers
Model:S7-3t & S8-3t Transesophageal (TEE) Transducers
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Ultrasound LLC
Issue:The manufacturer has recently discovered an issue associated with a small number of S7-3t and S8-3t Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/9/2019