Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:25036
Brand Name:Medifab Shuttle Discovery System ( Pushchair and Indoor base options with seating system)
Model:Verve Stroller
Affected:
Software version:
Recalling Organisation:Medifab Limited
1 Pereita Drive
Rolleston 7675
Contact Information:0800 543 343
Manufacturer:Medifab Limited
Issue:The manufacturer has identified that if the antivibration nut is overtightened, it will damage the T bolt within the folding joint which will lead to a greater gap at the folding joint. This could provide a finger entrapment for the user and anyone interacting with the frame, potentially leading to serious injury.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/9/2019