Detail of Recall

Type of Product:Device
Medsafe Reference:25001
Brand Name:Rocket Fetal Blood Sampling Kit
Model:
Affected:Multiple Lots
Software version:
Recalling Organisation:Rocket Medical (NZ) Ltd
Sidekick Auckland North Ltd
Level 1
445 Lake Road
Takapuna 0622
Contact Information:+64 021 198 8488
Manufacturer:Rocket Medical PLC
Issue:The manufacturer has received a complaint regarding the use of AXONGESIC ethyl chloride spray with Rocket FBS Amnioscopes. On contact with the spray, a white deposit was observed on the surface of the amnioscope and on the fetal scalp. This suggests an incompatibility between ethyl chloride and ABS.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/8/2019