Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24994 |
| Brand Name: | Shower Trolley: Carevo, Concerto |
| Model: | Concerto & Basic Shower Trolley |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Arjo New Zealand Limited 34 Vestey Drive Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 000 151 |
| Manufacturer: | ArjoHuntleigh AB |
| Issue: | The manufacturer has become aware that any of the 4 Safety Catch(es) may exhibit a mechanical failure (break in one of the arms of Safety Catch), which in certain circumstances may contribute to the Side Support unintentionally opening and may lead to a potential risk of patient fall, leading to possible injuries to patient and/or caregiver. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/8/2019 |