Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24992 |
| Brand Name: | Tina-quant Soluble Transferrin Receptor (sTFR) |
| Model: | cobas c systems (excl c111) |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Roche Diagnostics (NZ) Ltd 15 Rakino Way Mt Wellington AUCKLAND 1644 |
| Contact Information: | 0800 363 622 |
| Manufacturer: | Roche Diagnostics GmbH |
| Issue: | It was observed that the current claim for rheumatoid factors (RF) interference cannot be confirmed for the current STFR assay. Currently, the claim for this assay is that there is no significant interference of RF up to a concentration of 750 IU/mL for cobas c systems. The most recent internal investigations have confirmed that there is no significant interference of rheumatoid factors up to a concentration of 150 IU/mL for cobas c systems. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/8/2019 |