Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24982 |
| Brand Name: | Quantel - laser, diode |
| Model: | EASYRET LASER W/ HS STYLE S/L & TABLE |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Device Technologies New Zealand 47 Arrenway Drive Albany AUCKLAND 0632 |
| Contact Information: | 09 913 2000 |
| Manufacturer: | Quantel Medical |
| Issue: | During Easyret laser use, a software issue has been identified with the spot sensor of the device. When the single spot mode and 50 m spot size are selected, the power level delivered can be higher than the power level displayed on the laser screen. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/8/2019 |