Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:24952
Brand Name:Madsen Accuscreen
Model:Snap Leadwire
Affected:
Software version:
Recalling Organisation:Natus Medical New Zealand Limited
Alliott NZ Ltd
Level 2, 142 Broadway
Newmarket
Auckland 1149
Contact Information:0800 900 126
Manufacturer:GN Otometrics AS
Issue:It has been determined that the device does not fully meet the current regulatory standard for basic electrical safety and essential performance. There is a potential risk to the healthcare professional or patient of exposure to electrical shock from the electrode connector.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/8/2019
Hide menus
Show menus