Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24924 |
| Brand Name: | Infusion pump administration set_IIa |
| Model: | AS LVP 20D 3SS CV, AS LVP 20D LOW SORB, AS LVP 20D CV, AS LVP 20D DEHP 2VS, AS LVP 20D DEHP 2INJ, AS LVP 20D DEHP SS, AS LVP 20D 2SS CV, AS LVP 20D DEHP 2SS CV |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Becton Dickinson Ltd Unit 4 / 110 Mays Road Onehunga Auckland 1061 |
| Contact Information: | 0800 572 468 |
| Manufacturer: | BD Switzerland Sarl |
| Issue: | The manufacturer has identified an issue is caused by a variation in the wall thickness of the BD Alaris Pump Model 8100 Infusion Sets silicone pumping segment. There is potential for an incomplete occlusion leading to unintended delivery of medication when the pump is not running placed on pause with roller clamp opened or faster than expected drug delivery flow when the pump is infusing. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/8/2019 |