Detail of Recall

Type of Product:Device
Medsafe Reference:24906
Brand Name:Emotion 16
Model:11061628 (SOMATOM go.Up), 11061638 (SOMATOM go.All), 11061648 (SOMATOM go.Top)
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Shanghai Medical Equipment Ltd
Issue:The manufacturer has identified a potential issue concerning the injector holder falling off together with the injector and the storage box due to a mechanical defect in the Siemens Healthineers CT scanners. This potentially occurs when the injector and the storage box are being rotated horizontally around the injector arm by the operator.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/7/2019