Type of Product:Device
Medsafe Reference:24903
Brand Name:CRHF - Micra Transcatheter Pacing System MC1VR01
Model:Model number - MC1VR01
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:The manufacturer has received complaints involving the tether of the Micra Transcatheter Pacing System (TPS) becoming stuck upon tether retrieval. If resistance during tether removal is noted, this could result in the inability to remove the tether from the Micra Implantable Pulse Generator (IPG) and the acute dislodgement of the Micra IPG due to the increased force necessary for tether removal. The removal of the tether from the delivery system is a crucial step during implantation of the IPG. This stuck tether issue can potentially result in prolonged and possibly additional procedures, increasing harm to the patient.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Healthcare Professional
Date Commenced:31/7/2019