Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24896 |
| Brand Name: | Mammomat Fusion/Inspiration |
| Model: | MAMMOMAT Revelation VC10 with Biopsy Option |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare GmbH |
| Issue: | The manufacturer has become aware of an error that can occur when performing a biopsy using the InSpect function. The acquisition workstation can become unresponsive to normal user interaction. The issue can only occur in rare cases if all interactions happen: the workflow changes during biopsy by switching to the InSpect mode (the user double clicks the InSpect projection view icon) and, at the same time the user performs a right-mouse click at the acquisition workstation while the InSpect projection view is becoming active (changes color to white). |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 30/7/2019 |