Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:24889
Brand Name: ADVIA Centaur®, ADVIA Centaur® XP, ADVIA Centaur® XPT
Model:
Affected: Material Numbers – 10310308 & 10325366
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed the potential for the ADVIA Centaur Homocysteine assay (100 Test Kit SMN 10310374; 500 Test Kit SMN 10310375) to intermittently interfere with the Folate assay when testing is performed on the ADVIA Centaur, ADVIA Centaur XP or ADVIA Centaur XPT systems. The interference is observed intermittently with only the first replicate of Folate when the assay is tested immediately following Homocysteine assay testing.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/7/2019