| Type of Product: | Device |
| Medsafe Reference: | 24879 |
| Brand Name: | Intersurgical DOUBLE SWIVEL ELBOW and SUPERSET DOUBLE SWIVEL CATHETER MOUNT |
| Model: | Catalogue Numbers: IS-1898000 and IS-3521000 |
| Affected: | Batches: Batch: 2180346, 2180563, 2180731, 2181793, 2190292 and 2180397, 2180860, 2181474, 2181503, 2181704, 2182070, 2182119, 2190531, 2191217 |
| Software version: | |
| Recalling Organisation: | InterMed Medical Ltd 71 Apollo Drive Albany AUCKLAND 1311 |
| Contact Information: | 0800 333 444 |
| Manufacturer: | Intersurgical Ltd |
| Issue: | The manufacturer has received complaints about cracked ports in the swivel elbow where the flip cap is inserted, and these cracks leak if they reach a significant size. The cracks were not present at the time of manufacture but have occurred later on during storage. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 26/7/2019 |