Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24876 |
| Brand Name: | Radiometer ABL90Flex; ABL80Flex; ABL9 analysers |
| Model: | Serial numbers below 090R0565N003 that have an operating system version RXPE-UIM2743 v1.3.xxxxxx or version RXPE-UIM910 v1.3.xxxxxx |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Issue: | On May 14th 2019 Microsoft released a patch for a critical Remote Code Execution vulnerability in Remote Desktop Services (CVE-2019-0708). This vulnerability can be exploited remotely without authentication on systems that use Remote Desktop Services for Windows XP, Windows 7, Windows Server 2003 and Windows Server 2008 operating systems. RADIOMETER uses RXPE and RWES7, which are embedded versions of the Windows XP and Windows 7 operating systems customized for our applications. These versions have limited functionality compared to the full versions for which reason the vulnerability is limited to the version stated under affected product. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/7/2019 |