Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:24876
Brand Name:Radiometer ABL90Flex; ABL80Flex; ABL9 analysers
Model:Serial numbers below 090R0565N003 that have an operating system version RXPE-UIM2743 v1.3.xxxxxx or version RXPE-UIM910 v1.3.xxxxxx
Affected:
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:On May 14th 2019 Microsoft released a patch for a critical Remote Code Execution vulnerability in Remote Desktop Services (CVE-2019-0708). This vulnerability can be exploited remotely without authentication on systems that use Remote Desktop Services for Windows XP, Windows 7, Windows Server 2003 and Windows Server 2008 operating systems.
RADIOMETER uses RXPE and RWES7, which are embedded versions of the Windows XP and Windows 7 operating systems customized for our applications. These versions have limited functionality compared to the full versions for which reason the vulnerability is limited to the version stated under affected product.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/7/2019
Hide menus