Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24875 |
| Brand Name: | Medical Intelligence HexaPOD evo RT System |
| Model: | |
| Affected: | iGUIDE software versions 2.2.0 - 2.2.2 |
| Software version: | |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Medical Intelligence Medizintechnik GmbH |
| Issue: | Sometimes, although the HexaPOD has reached the target position (within the specified positioning accuracy), iGUIDE may incorrectly demand a verification scan. This behavior only occurs if the HexaPOD alone cannot reach the target and requires support from the Precise Treatment Table. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Wholesale |
| Date Commenced: | 22/7/2019 |