Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24865 |
| Brand Name: | Cleveland CT |
| Model: | the Philips Ingenuity CT, iCT, Brilliance 64, IQon Spectral CT, and Ingenuity TF PET/CT |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems |
| Brand Name: | Ingenuity TF PET/CT |
| Model: | Philips Ingenuity CT, iCT, Brilliance 64, IQon Spectral CT, and Ingenuity TF PET/CT |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems |
| Issue: | The manufacturer has identified that in certain instances when a patient exhibits an arrhythmia or varying heart rate during a Step and Shoot acquisition, images from the acquisition following the heart rate variation, may be labeled in the incorrect phase. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/7/2019 |