Detail of Recall

Type of Product:Device
Medsafe Reference:24819
Brand Name:EVOTECH Integrated Endoscope Disinfection System
Model:Serial Number: 5042100008
Affected:
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Advanced Sterilization Products Inc
Issue:The manufacturer has recently determined that some EVOTECH® ECR systems equipped with certain printers exceeded the radio frequency (RF) emission specified by IEC-60601-1-2 and CISPER 11 Class A for radio frequency disturbance.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/7/2019