Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24819 |
| Brand Name: | EVOTECH Integrated Endoscope Disinfection System |
| Model: | Serial Number: 5042100008 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Johnson & Johnson (New Zealand) Limited 507 Mt Wellington Highway Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 922 638 |
| Manufacturer: | Advanced Sterilization Products Inc |
| Issue: | The manufacturer has recently determined that some EVOTECH® ECR systems equipped with certain printers exceeded the radio frequency (RF) emission specified by IEC-60601-1-2 and CISPER 11 Class A for radio frequency disturbance. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/7/2019 |