Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24812 |
| Brand Name: | Interpretive multichannel electrocardiograph |
| Model: | Affected Product: All MAC VU360 Systems P/N: 2030360-001 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems Information Technologies Inc |
| Issue: | The manufacturer has become aware of the MAC VU360 system may intermittently display an incorrect patient ID or visit number on the screen after scanning the patient's barcode or may have the incorrect patient demographics appear on the patient banner. This may result in the ECG report being assigned to the incorrect patient. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/7/2019 |