Detail of Recall

Type of Product:Device
Medsafe Reference:24796
Brand Name:AQUIOS IMMUNO-TROL Low Cells
Model:
Affected:Lots - 6180091 (6180091K) expiry 2019-12-07
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has confirmed through internal testing of this product lot that the system may include a significant portion of the monocyte population in the lymphocyte region during automated data analysis due to insufficient separation of the two populations in the side scatter vs. CD45 plot.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/7/2019