Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:24771
Brand Name:Ellipse DR CD2377-36QC
Model:models CD2377-36QC
Affected:
Software version:
Recalling Organisation:Abbott Medical New Zealand Limited
4 Pacific Rise
Mt Wellingtion
Auckland 1640
Contact Information:0800 756 269
Manufacturer:St Jude Medical (USA)
Issue:Abbott Ellipse ICDs utilize aluminum wires to connect the high voltage components of the device, and the electronics are then encapsulated in epoxy. During final manufacturing testing, two electrical failures were identified in a limited lot of manufactured devices due to damaged aluminum wires.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Healthcare Professional
Date Commenced:26/6/2019
Hide menus