Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:24740
Brand Name:Tina-quant Myoglobin Gen.2 imprecision
Model:
Affected:Tina-quant Myoglobin Gen.2 (MYO2) assay on cobas c 311/501/502 with reagent lot 34986001
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:Internal investigations have confirmed QC imprecision and calibration failures due to Dup E. It was shown that in the affected measurements a drop in the signal in the reaction kinetics occurred. This drop is most likely caused by a break-up of a precipitate triggered by mixing in a neighbouring cell. As a consequence, a decreased sample recovery is generated.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/6/2019