Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24706 |
| Brand Name: | DiaClon Rh-Subgroups + K |
| Model: | |
| Affected: | All lot numbers until further notice |
| Software version: | |
| Recalling Organisation: | Bio-Rad New Zealand Level 30 48 Shortland Street AUCKLAND 1010 |
| Contact Information: | 0508 805 500 |
| Manufacturer: | Diamed NZ div of DiaMed Australia Pty Ltd |
| Issue: | The manufacturer has been able to confirm that interference can be observed in the anti-e microtube with QC samples antigen e negative (RH:-5) This interference is giving a pinkish colour in the anti-e microtube with QC samples antigen e negative (RH:-5), the instruments may read it as wR, wF, ? or in extreme case as +. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/6/2019 |