Type of Product:Device
Medsafe Reference:24706
Brand Name:DiaClon Rh-Subgroups + K
Model:
Affected:All lot numbers until further notice
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Diamed NZ div of DiaMed Australia Pty Ltd
Issue:The manufacturer has been able to confirm that interference can be observed in the anti-e microtube with QC samples antigen e negative (RH:-5)
This interference is giving a pinkish colour in the anti-e microtube with QC samples antigen e negative (RH:-5), the instruments may read it as wR, wF, ? or in extreme case as +.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/6/2019