Type of Product:Device
Medsafe Reference:24686
Brand Name:Alinity ci-series System Control Module
Model:
Affected:All Lots
Software version:
Recalling Organisation:Abbott Diagnostics Division
Ground Floor, Building D
4 Pacific Rise
Mt Wellington
AUCKLAND 1640
Contact Information:0800 656 233
Manufacturer:Abbott Diagnostics Division
Issue:The manufacturer has identified an issue with all on-market versions of Alinity ci-series Software where re-use of reaction vessels (RVs) may occur after a system Stop. This issue only occurs if the system is transitioned from Processing to Stopped to Idle and has the potential to cause incorrect results.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/6/2019