Detail of Recall

Type of Product:Device
Medsafe Reference:24685
Brand Name:Artis Q, Q.zen, zeego, zee systems
Model: Artis Q floor-10848280, Artis Q ceiling-10848281, Artis Q zeego-10848283, Artis Q biplane-10848282, Artis Q.zen biplane-10848355, Artis zee ceiling-10094137, Artis zee floor-10094135, Artis zee biplane-10094141, Artis zeego-10280959
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:On the DCS Extended display suspension system, it is possible for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off. The problem occurs sporadically and is dependent on the frequency of use of the monitor suspension system.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/6/2019