Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 24664 |
| Brand Name: | X-Ray System, Mammographic, Stationary, Digital |
| Model: | All Senographe Pristina systems that are equipped to receive a biopsy device. |
| Affected: | |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems SCS |
| Issue: | The manufacturer has identified that if the Senographe Pristina gantry is rotated and the biopsy positioner is attached, the angulation angle may unexpectedly change during a motorized angulation motion, due to insufficient brake performance. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/6/2019 |