Detail of Recall

Type of Product:Device
Medsafe Reference:24654
Brand Name:Medical device electrical cable, general purpose
Model:ECG trunk cables and leadwires.
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Healthcare Finland Oy
Issue:The manufacturer has identified that the ECG trunk cables and leadwires may reduce the amount of energy reaching the patient during defibrillation, potentially limiting success of defibrillating the patient's rhythm. If this issue occurs during a defibrillation event, it may not be noticeable to the caregiver and could contribute to an adverse patient outcome.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/5/2019